Search
Home › Products ›
CT6904A/CT6904A-1

Hioki CT6904A/CT6904A-1

Hioki
CT6904A / CT6904A-1 AC/DC ,4 MHz/500A, 32mm
Best-in-class Measurement Bandwidth with High Accuracy

Product Description

Hioki's pass-through type AC/DC current sensors can be connected directly to Hioki Power Analyzers and Memory HiCorders. Also, it can be connected to Power Meters, Oscilloscopes, DMMs, etc. by using an optional sensor unit.

The CT6904A has a frequency bandwidth of DC to 4 MHz (Model CT6904A-1 is DC to 2 MHz) and a rated current of 500 A. It is used in research and development fields such as loss measurement of high-frequency reactors, wireless charging systems, and EV inverter motors.

CAT III 1000 V

CT6904A: Cable length: 3 m (9.84 ft.)

CT6904A-1: Build-to-order, cable length: 10 m (32.81 ft.)

Unparalleled technology driving the evolution of current measurement

High-accuracy sensors uses the "zero flux method (flux gate detection type)." High-frequency currents are detected with the winding (CT method), and DC to low frequency currents are detected using a "flux gate."

Wide frequency band

The CT6904A has a frequency bandwidth of DC to 4 MHz (Model CT6904A-1 is DC to 2 MHz), which is top-class for a 500 A rated pass-through sensor. In combination with the Power Analyzer PW8001, high frequency and low power factor can be measured accurately.

High Noise Resistance

By completely shielding the oppositely arranged split coils with a unique-shaped solid shield, it achieves both wide bandwidth and superior noise resistance. Accurate measurement can be performed without being affected by ambient voltage.

Technical Specifications

Key Features:

  • 500 A (rms) rated for measurement of large currents
  • CT6904A: DC to 4 MHz (±3 dB), CT6904A-1: DC to 2 MHz (±3 dB) wide measurement frequency range
  • ±5 ppm excellent linearity
  • 120 dB (100 kHz) high Common-Mode Rejection Ratio (CMRR)
  • Guaranteed combined accuracy in combination with Hioki Power Analyzers

Specifications:

Rated current500 A AC/DC
Max. allowable inputWithin the derating range, However, up to ±1000 A peak is allowable if within 20 ms. (design value)
Frequency bandwidthAmplitude: DC to 4 MHz (CT6904A), DC to 2 MHz (CT6904A-1)
Phase: DC to 1 MHz
Linearity±5 ppm Typical (23°C (73°F))
Offset voltage±10 ppm Typical (23°C (73°F), no input)
Basic accuracyDC (Amplitude: ±0.025 % rdg. ±0.007 % f.s., No phase specification)
45 Hz ≤ f ≤ 65 Hz (Amplitude: ±0.02 % rdg. ±0.007 % f.s., Phase: ±0.08°)
Output voltage rate4 mV / A rated (This device outputs AC+DC voltage via the Sensor Unit)
Max. rated voltage to earth1000 V CAT III
Core diameterφ 32 mm (1.26 in.)
Operating temperature, humidity-10°C to 50°C (14°F to 122°F)
80% RH or less (with no condensation)
Power supplyPower supplied from compatible Hioki power analyzers, sensor units, or current units.
Max. rated power7 VA Max. (500 A/55 Hz measurement, with a power supply of ±12 V)
Dimensions and mass139 mm (5.47 in.) W × 120 mm (4.72 in.) H × 52 mm (2.05 in.) D
CT6904A: 1.05 kg (37 oz.), cable length 3 m (9.84 ft.)
CT6904A-1: 1.35 kg (47.6 oz.), cable length 10 m (32.81 ft.)
Included accessoriesInstruction Manual ×1, Carrying case ×1, Color labels (for channel identification) ×1, Operating Precautions ×1
Product Resources

Long-term Logging & Feature Highlights

Directly record to SD cards or internal memory for weeks of continuous monitoring without data loss. [Placeholder text box for custom product feature highlights populated later]

Official Australian Service & Calibration Partner

Calibration, Servicing & Warranty Support

Power Parameters is the official authorized Australian service and technical support facility. We provide traceable calibration, certified warranty repairs, scheduled maintenance, and genuine OEM parts from our climate-controlled laboratory in Westmeadows, Victoria, as well as on-site calibration across Australia.

Fast turnaround • Certificate included
Traceable Calibration Certificates
Genuine OEM Parts & Diagnostics
Fast 3–5 Day Turnaround
In-House Lab (VIC) & On-Site