Hioki clamp meters are best-in-class measuring devices incorporating Hioki’s expertise in current sensing technology. The CM4373-50 is a 2000 A True RMS clamp meter that features stronger jaws, IP54 dustproof and waterproof enclosure to enable professionals to work quickly in tough situations including the development, repair, and maintenance of equipment and devices such as electric vehicles. Additionally, the high-voltage DC measurement capability of up to 2000 V (*1) is ideal for the clamp maintenance inspections of electrical and solar power equipment.
CAT IV 600 V, CAT III 1000 V
*1: When using P2000: CAT III 2000 V, CAT IV 1000 V
When the Wireless Adapter Z3210 is installed, Bluetooth® wireless communication is enabled. It transfers measurements to your smartphone or tablet and allows you to quickly create reports with field photos and drawings.
When combined with GENNECT Cross, you can perform simple harmonic analysis. Applications include problem analysis of power supply systems.
Problems that can be caused by harmonics
The clamp meters can simultaneously measure inrush current in RMS as well as maximum crest values at motor startup and for welding currents. The clamp meters automatically detect the duration of the inrush current (which can range from several dozen milliseconds to several hundred milliseconds) and measure the current during that interval, enabling them to yield more accurate measurements than standard clamp-on meters whose measurement interval is fixed to 100 ms.
| DC Current range | 600.0 A/2000 A, Basic accuracy: ±1.3% rdg. ±0.3 A (600 A range) | |||||
|---|---|---|---|---|---|---|
| AC Current range | 600.0 A/2000 A (10 Hz to 1 kHz, True RMS), Basic accuracy 45 – 66 Hz: ±1.3% rdg. ±0.3 A (at 600 A) | |||||
| Crest factor | 600.0 A range: 3 or less 2000 A range: 2.84 or less | |||||
| DC+AC Current range | 600.0 A/2000 A (10 Hz to 1 kHz, True RMS), Basic accuracy DC, 45-66 Hz: ±1.3% rdg. ±1.3 A (at 600 A) | |||||
| DC Voltage range | 600.0 mV to 1000 V (When using P2010: 600.0 V to 2000 V) | |||||
| AC Voltage range | 6.000 V to 1000 V, 4 ranges (15 Hz to 1 kHz, True RMS), Basic accuracy 45 – 66 Hz: ±0.9% rdg. ±0.003 V (at 6 V) | |||||
| DC+AC Voltage range | 6.000 V to 1000 V, 4 ranges, Basic accuracy DC, 45 – 66 Hz: ±1.0% rdg. ±0.013 V (at 6 V) | |||||
| Resistance range | 600.0 Ω to 6.000 MΩ, 5 ranges, Basic accuracy: ±0.7% rdg. ±0.5 Ω (at 600 Ω) | |||||
| Capacitance range | 1.000 μF to 1000 μF, 4 ranges, Basic accuracy: ±1.9% rdg. ±0.005 μF (at 1 μF) | |||||
| Frequency range | 9.999 Hz to 999.9 Hz, 3 ranges, Basic accuracy: ±0.1% rdg. ±0.003 Hz (at 9.999 Hz) | |||||
| Temperature (K) | -40.0 to 400.0 ˚C, add temperature probe accuracy to basic accuracy of ±0.5% rdg. ±3.0 ˚C | |||||
| Voltage detection | Hi: 40 V to 600 V AC, Lo: 80 V to 600 V AC, 50/60 Hz | |||||
| Harmonic levels | [With Z3210 installed] Voltage/Current harmonic levels up to 30th, Content factor, Total harmonic distortion ratio *Harmonics can be displayed with our free app GENNECT Cross. | |||||
| Other functions | DC power, Continuity check, Diode check, Automatic AC/DC detection, Pass/fail judgement function of DC A and DC V, Max/Min/Average/PEAK MAX/ PEAK MIN value display, Low-pass filter function, Display value hold, Auto hold, Back light, Auto-power save, Buzzer sounds, Zero-adjustment, etc. | |||||
| Dustproof, waterproof | IP20 (Voltage measurement in a completely dry condition. When jaw closes) IP54 (While in storage) | |||||
| Power supply | LR03 Alkaline battery ×2 Continuous use: 40 hr (without Z3210 installed), 24 hr. (with Z3210 installed and using wireless communications) Other conditions: 100 A AC measurement, backlight off, 23°C reference value | |||||
| Core jaw diameter | φ55 mm (2.17 in), Jaw dimension: 92 mm (3.62 in) W×18 mm (0.71 in) D | |||||
| Dimensions and mass | 65 mm (2.56 in) W×250 mm (9.84 in) H×35 mm (1.38 in) D mm, 530 g (18.7 oz) | |||||
| Included accessories | Test Lead L9300, Carrying Case C0203, LR03 Alkaline battery ×2, Instruction Manual×2, Operating Precautions ×1 | |||||
Directly record to SD cards or internal memory for weeks of continuous monitoring without data loss. [Placeholder text box for custom product feature highlights populated later]
Power Parameters is the official authorized Australian service and technical support facility. We provide traceable calibration, certified warranty repairs, scheduled maintenance, and genuine OEM parts from our climate-controlled laboratory in Westmeadows, Victoria, as well as on-site calibration across Australia.

Model: CM4373-50